The use of computerized intermediate frequency therapy instrument

Detailed use is as follows:

1, the power cord on the net power connector into the electrotherapy machine equipment power input socket, plug into the fixed net power socket.

2, the electrode plate will be placed in the patient's affected area, the electrode plate and the human body should be used between the dry towel as a mat, the thickness of which is based on the treatment site, generally 1-2 cm. the mat the more after the deeper tissues of the thermal effect of the better. The affected part should be placed between the two electrodes as far as possible when using the device. To prevent electrode displacement, the electrode and pad can be tied to the affected part with a dry bandage.

3. Select the "treatment time" (timing time) according to your treatment needs.

4. Turn the "Output Selection" to "Preheat".

5, the "power switch" from "off" to "on", "switch" green light, the machine power supply is on, the electronic tube is turned on, the electronic tube (timing time). The green light of the "switch" will be on, and the power supply in the machine will be turned on, and the electronic tube will be lighted to start the "preheating".

6, the "power supply voltage adjustment" screw to the voltmeter indicates 220V. "Preheat" for two minutes.

7, the "output selection" rotary to "treatment 1" file, then the yellow light, the machine began to work normally. The ammeter indicates the size of the oscillating tube screen flow.

8, turn the "output tuning", the ammeter indication also changes, when the resonance point is reached, the ammeter indication of the maximum. Treatment is best done at the resonance point.

9, according to the needs of the treatment, you can choose different values of current. Such as "output selection" of the "1" file is not enough dose, you can choose "2", "3", "4" file. "4" gear to increase the dose. Each time you change gears, you need to re-tune it so that it reaches resonance.

10, when the treatment time, the output is automatically cut off, the ammeter indicates zero, should be "output selection" in order to return to the "preheating", turn off the power supply, and then remove the electrode plate from the affected part of the patient.

The role of computerized intermediate frequency therapeutic instrument:

Computerized intermediate frequency therapeutic instrument can be used to treat cervical spondylosis, arthralgia, osteomalacia, sciatica, periapical, functional point stimulation, analgesia, soft tissue injury, anti-inflammatory, decongestive and other common diseases.