What are the risks of cooling the sterilizer and opening the room for air conditioning and refrigeration?

Requirements of high temperature steam sterilizer for cold air discharge.

1, non-automatic temperature control sterilizers are mostly small sterilizers: for such sterilizers, cold air must be discharged. If the cold air is not discharged, it will contribute to the working pressure of the pressure cooker after thermal expansion. At this time, the pressure and temperature can't correspond to each other one by one, and the displayed temperature is higher than the actual working temperature in the pot. The working temperature of sterilizer can not reach the set sterilization temperature, which will affect the sterilization effect.

2. Automatic temperature-controlled sterilization pot: The automatic temperature-controlled sterilization pot monitors the temperature change in the pot through the heat-sensitive electrode, and the temperature displayed on the display screen is the real working temperature in the pot, which can ensure the sterilization temperature requirement even if the cold air is not exhausted. Due to the sterilization effect of saturated steam *, cold air should also be discharged when using the automatic temperature control sterilizer to improve the sterilization effect. Generally, there is a drain and exhaust valve under the pot, which can be rotated a little to the left (drain water to the right) to allow a small amount of steam to be discharged. Some sterilization equipment is equipped with an adjustable exhaust valve, which can always be slightly opened during sterilization and discharge cold air.

The intelligent sterilization operation program of the autoclave ensures the sterilization effect, improves the working efficiency, enhances the safety performance, and realizes the control system that the sterilization program is completed by itself and the whole sterilization operation process is unattended. The steam sterilization of automatic autoclave has reached the design concept of surpassing similar products in Europe and America. With superior performance and perfect configuration, it meets the needs of medical and health care, universities, biological research, food and chemical industry and other units.