1 Maintenance of the equipment first dust, with a brush and vacuum cleaner to remove the dust.
2 Components or connectors found to have a loose phenomenon, to be immediately tightened; there are components rust, if necessary, to replace.
3 Regularly check the following parts:
3.1 Check the power indicator above the keyboard and channel, the indicator light is yellow when the device is powered on.
3.2 Check the X-ray indicator above the keyboard and channel The indicator light is on in red when the rays are emitted.
3.3 Check the emergency stop buttons above the keypad and channels. Pressing any of the emergency stop buttons immediately powers down the unit. When the buttons are not reset, the device cannot start again.
3.4 Check all the keys on the keyboard and their corresponding functions, press the zoom key, the image can be zoomed; press the image processing key, the displayed image can make the corresponding transformation.
3.5 Check the exit and entrance light barriers. Put an opaque object on the running conveyor belt, the ray can be opened normally. When the equipment is powered off, wipe the lenses of the light barriers and the glass mounted on the channel walls with a dry silk cloth.
3.6 Check the motorized pulley. Press the appropriate button, the drum motor can run in the intended direction. The noise of the drum is normal and there is no oil leakage.
3.7 Check the travel switches on the probe box and collimator. Removing any of the covers, the rays were able to stop firing.
3.8 Check for conveyor belt deflection. Check to see if the conveyor belt has deflected relative to the motor.
3.9 Check the cooling fan. Check that the cooling fan is working properly and remove dust from the guard and bracket.