Procurement requirements and technical requirements
I. Procurement requirements and detailed technical requirements
(A) Each terminal of each floor is required to be prepared:
1. Specific requirements:
General wards are set up for each bed with one oxygen terminal and one suction terminal; rescue rooms and trauma rooms are set up for each bed with two oxygen terminals and two suction terminals; each bed is equipped with a set of infusion hangers (straight rails for general wards, straight rails for trauma rooms, straight rails for trauma rooms, straight rails for trauma rooms, straight rails for trauma rooms). Each bed is equipped with one set of infusion hanger (straight rail for general ward, U-rail for trauma room, rescue room and labor and delivery room).
2, the number of terminals:
The second floor: 12 wards (including 10 rooms for 3 people, 1 room for 2 people and 1 room for 1 person), 33 beds, 33 oxygen terminals and 33 suction terminals.
1 resuscitation room, 1 bed, 2 oxygen terminals, 2 suction terminals.
1 waiting room, 3 beds, 3 oxygen terminals, 3 suction terminals.
2 labor rooms, 4 beds, 4 oxygen terminals, 4 suction terminals.
***Total: 42 oxygen terminals, 42 suction terminals
Third and fourth floors: 13 wards (including 10 3-bed rooms, 1 2-bed room and 2 1-bed rooms), 34 beds, 34 oxygen terminals and 34 suction terminals.
Resuscitation room 1, 1 bed, 2 oxygen terminals, 2 suction terminals.
1 room for debridement, 1 bed, 2 oxygen terminals, 2 suction terminals.
***Total: 76 oxygen terminals, 76 suction terminals.
Fifth, sixth and seventh floors: 13 wards (including 10 rooms for 3 people, 1 room for 2 people and 2 rooms for 1 person), 34 beds, 34 oxygen terminals and 34 suction terminals.
One resuscitation room, 1 bed, 2 oxygen terminals, 2 suction terminals.
***Total: 108 oxygen terminals, 108 suction terminals.
Tenth floor: 6 operating rooms, 12 oxygen terminals, 12 suction terminals.
Total: 87 wards, 218 beds, 238 oxygen terminals, 238 suction terminals, 218 infusion hangers (including 18 U-rail hangers and 200 straight-rail hangers).
(2) piping system technical requirements
1, oxygen piping and connectors are all made of stainless steel, material OCr18N19 (domestic famous brand).
2, oxygen main pipe, riser diameter of Φ24 × 2 stainless steel pipe.
3, oxygen horizontal pipe, pipe diameter: Φ14 × 1 stainless steel.
4, into the wards and equipment within the belt branch pipe: Φ8 × 1 stainless steel pipe. (Resuscitation room, ICU room branch pipe for Φ10 × 1 stainless steel pipe).
5. The main pipe of the absorption system building and the cross pipe of the ward are made of international galvanized steel pipe, and the branch pipe and connecting parts into the ward are OCr118N19 stainless steel pipe. (Domestic famous brand).
6, absorption of the main pipe, riser pipe diameter: DN80 galvanized steel pipe.
7, attracting cases of cross-pipe diameter: DN40 galvanized steel pipe.
8, attraction ward room and equipment with branch pipe: Φ10 × 1 stainless steel pipe. (Resuscitation room, ICU room tube for Φ14 × 1 stainless steel tube).
(C) ward and ward equipment technical requirements
1, oxygen system end-use pressure of 0.4Mpa (adjustable), attraction system using pressure: -0.03 Mpa-0.07 Mpa (adjustable).
2. A set of oxygen secondary decompression box is installed in each ward, and the device in the decompression box adopts dual-channel design with one for use and one for backup.
3. A set of oxygen flow meter is installed in each ward (digital display, accurate measurement).
4, each ward oxygen suction cross-pipe design has a set of maintenance valves.
5. The equipment belt must be made of national standard aluminum with thickness not less than 2㎜. (Domestic famous brand)