Which biochemistry analyzer performs better, the Myriad BS-300 BS380 BS400? Why?

I recommend the BS390, the model is the bidding version of the BS380, if the bidding if the price can be taken lower, the performance is exactly the same, the difference between the three models is mainly as follows:

Gap one, BS300 is a maximum of 300 items/hour test speed, 380 and 400 is a constant speed of 380, 400 items per hour test,

Gap two, BS300 reaction cups use disposable reaction cups, the use of robots to automatically replace the reaction cups associated with the BS380/400 use automatic cleaning device to repeatedly clean the use of reaction cups, but to a certain extent after the reaction cups should also be replaced, the main consumption of concentrated cleaning fluid

380 and 400 gap is mainly in the grade, 380 concentrated cleaning fluid needs to be manually diluted to add. And because there is no use of large air pumps, so the discharge tube needs to be self-flowing discharge, 400 degree of automation is higher, the use of large air pumps, concentrated cleaning fluid are added to the original liquid machine automatically dilute the use of

Characteristics of the introduction of the end of the 300 with the price difference between the 390 is not very large, a look at the understanding of the cost-effectiveness of the 380 is the highest, the 300 is basically the elimination of the way of working, do not! Recommended