Minimum disinfection time and temperature of peroxyacetic acid for reuse of dialyzers

The minimum disinfection time of peroxyacetic acid for dialyzer reuse is 30 minutes and the temperature is 25℃.

1. In the process of dialyzer reuse, it is necessary to select the appropriate disinfectant and disinfection conditions in order to ensure that the dialyzer is thoroughly disinfected.

2, The shortest disinfection time of peroxyacetic acid at 25℃ is 30 minutes, which can effectively kill the microorganisms on the surface of dialyzer.

3. Insufficient disinfection time or low temperature may result in residual microorganisms on the surface of the dialyzer, thereby increasing the risk of infection.

4. Therefore, the minimum disinfection time of peroxyacetic acid for reuse of dialyzers is 30 minutes and the temperature is 25°C, aiming to ensure thorough disinfection of the surface of dialyzers to safeguard the health and safety of dialysis patients. Dialyzers are highly sensitive medical devices that require thorough disinfection to avoid cross-infection. Peroxyacetic acid is a widely used disinfectant that effectively kills microorganisms such as bacteria, viruses and fungi.